1700 Series
PRINT E-MAIL BOOKMARK  

Veriteq 1700 Series thermocouple data loggers are miniature stand-alone devices developed for in-field recording of Type J, K, S, T, E, and R thermocouple temperature probes. Designed for long-term operation in harsh industrial and field applications, these data loggers are equipped with automatic cold junction temperature compensation. The units feature a high degree of immunity from electrical interference, physical shock and vibration allowing them to be mounted on or near large electrically-powered equipment or systems. The non-volatile EEPROM memory ensures long-term data protection in the event of damage to the unit.

Veriteq software is required to setup the data logger and download data collected.


  ENQUIRY
MODELS

Product description Code
VL-1700 tamper-proof logger for the Pharmaceutical/Biotech industry with 1 channel using Type T thermocouple.
Software to be used: vLog
VVL170021N
VL-1700 tamper-proof logger for the Pharmaceutical/Biotech industry with 4 channels, each using Type T thermocouple.
Software to be used: vLog
VVL170051N
SP-1700 logger with 1 channel, narrow range (-220°C to +210°C)
(No thermocouples included) Software to be used: Spectrum
VSP170021N
SP-1700 logger with 1 channel, wide range (-100°C to +1760°C)
(No thermocouples included) Software to be used: Spectrum
VSP170021W
SP-1700 logger with 4 channels, narrow range (-220°C to +210°C) (No thermocouples included) Software to be used: Spectrum VSP170051N
SP-1700 logger with 4 channels, wide range (-100°C to +1760°C)
(No thermocouples included) Software to be used: Spectrum
VSP170051W

 


MORE INFORMATION

FEATURES

Product details for 1700 SERIES
bullet Single & multi-input data logger models record temperature from -240°C
  to 1760°C
Productseperator
bullet User-configurable data logger inputs accept Type J, K, T, E, R and S
  thermocouples
Productseperator
bullet High impedance inputs maximize accuracy and allow use with long or thin
  gauge thermocouple probes
Productseperator
bullet Completely self-contained with internal memory and 10-year battery
Productseperator
bullet 70,000 sample memory
Productseperator
bullet Two year limited warranty
Productseperator
bullet Automatic cold junction temperature compensation

SPECIFICATIONS

Product specifications for 1700 SERIES
Measurement range Type K: -220°C to +210°C (Narrow range)
-100°C to +1370°C (Wide range)
Type J: -130°C to +160°C (Narrow range)
-50°C to +900°C (Wide range)
Type T: -240°C to +190°C (Narrow range)
-100°C to +350°C (Wide range)
Type E: -110°C to +140°C (Narrow range)
-50°C to +740°C (Wide range)
Type R: -50°C to +790°C (Narrow range)
-50°C to +1760°C (Wide range)
Type S: -50°C to +850°C (Narrow range)
-50°C to +1700°C (Wide range)
Resolution @ mid-range Type K: 0.10°C (Narrow range) 0.37°C (Wide range)
Type J: 0.08°C (Narrow range) 0.29°C (Wide range)
Type T: 0.11°C (Narrow range) 0.34°C (Wide range)
Type E: 0.07°C (Narrow range) 0.20°C (Wide range)
Type R: 0.40°C (Narrow range) 1.3°C (Wide range)
Type S: 0.43°C (Narrow range) 1.5°C (Wide range)
Clock accuracy ±1 min/month at 0–50°C
Memory size 70,000 data points
Memory protection Data retention > 20 years backup without power
Memory modes User-selectable. Wrap (FIFO) or stop when memory is full. User selectable start/stop time
Data setup and offload VL1700 series data logger is compatible with vLog & viewLinc (for on-going monitoring & alarming). SP1700 series data logger is compatible with the following software packages: Spectrum, viewLinc (for on-going monitoring & alarming), Veriteq sGo (for Palm OS), Spectrum Excel Add-in and Spectrum API (for custom applications)
Data communication RS-232 serial port, Ethernet, PoE
Power source Internal Lithium battery with life of 10 years at 1 minute sampling rate
Enclosure Injection-molded beige ABS plastic
Dimensions 71 x 53 x 18mm
Weight 60gm

DOWNLOADS

Documentation and software downloads for 1700 SERIES
Data Sheet 1700 Series

ACCESSORIES

Recommended accessories for 1700 SERIES
Product description Code
vLog Version 4.21 software with USB cable VSWUSBD421
IQ/OQ validation protocol for vLog software 4.21 VDOCVP6
Spectrum software with USB cable VSWSPUSBD
NEMA 4X (IP66) rated enclosure VACENC01
USB cable VUSBDB9
Serial cable VDB96
© 2008 Easy Solutions, All rights reserved.