Anville 825 Thermal Validation Logger
PRINT E-MAIL BOOKMARK  

Anville 825 recorder has been designed to meet the requirements of autoclave and tunnel validation in the Pharmaceutical industry. It has an accuracy of ±0.25°C which is a worst case "out of the box" value including all system errors but excluding thermocouple sensor error.

Some of the design features that separates it from the majority of the loggers include using correctly compensated miniature type T thermocouple connectors with matching class 1 thermocouple wiring to the internal cold junction sensor, a 1/3 DIN Pt100 sensor that measures the cold junction temperature to an accuracy of ±0.1°C at 0°C, design of the physical layout of the cold junction which reduces any thermal scatter to a typical deviation of ±0.05°C and the usage of individual high speed amplification and analogue to digital conversion on each input channel with conversion to engineering units (°C/°F) carried out by the microprocessor in the device.

The logger is provided as a part of the thermal validation system that includes Eazival SE software, reference thermometer, calibration bath, pressure transducer and pneumatic calibrator. The software complies with EN554, EN285, HTM2010, HTM2030 and 21CFR Part 11.


  ENQUIRY
MODELS

Product description Code
Anville 825 Thermal Validation Logger EZ825

 


MORE INFORMATION

FEATURES

Product Details for 825 Series Data Logger
bullet Accepts up to 16 T-type thermocouples, 2 4-20mA transducer inputs
  and an event marker
Productseperator
bullet Modular. Multiple 825 Series can be connected to the same software
Productseperator
bullet Over -50°C to 300°C measurement range, the overall system accuracy
  is ±0.25°C
Productseperator
bullet Combined uncertainties of the complete measurement system,
  including sensors is 0.41°C (UKAS 95% confidence)
Productseperator
bullet Works with EaziVal software provided free of cost. Compatible
  with TQSoft software
Productseperator
bullet Eazival allows unlimited number of logger setup configurations
  to be saved for future use.
Productseperator
bullet Calibration equipment inventory details the data logger, temperature bath,
  temperature reference, pressure transducer and pressure calibrator. The formal calibration details for each piece of equipment include key calibration dates and calibration certificate numbers.
Productseperator
bullet All calibration results are stored in tamper-proof encrypted files
Productseperator
bullet Calibration report lists the equipment used along with the actual
  calibration results for each channel showing the error at each of the three set-points, the set-point value and the error after calibration correction.
Productseperator
bullet Cycle information window shows immediately whether all the critical
  sterilisation parameters for the machine under test have been satisfactorily achieved
Productseperator
bullet Software automatically organises all the files created by the system so
  that all thermometric data (chart and data listing), reports, calibrations and paperwork are immediately available by selecting the appropriate machine in the database
Productseperator
bullet "Paperwork" button in the toolbar brings up a wide choice of appropriate
  test sheets and templates for use when carrying out weekly, quarterly or yearly tests on porous-load and bench top sterilisers

SPECIFICATIONS

Product specifications for 825 Series Data Logger
Number of input channels 18 (16 Type T or Type K thermocouple inputs, 2 4-20mA inputs)
Analog input terminal shape Miniature type T thermocouple connectors
Scanning speed All channels in less than 1 sec.
Thermocouple Type Type T or Type K thermocouple
Measurement range - Temperature -200 to 400°C
Measurement accuracy - Temperature ±0.25°C (-50.0°C to 300.0°C); ±1.0°C (-200.0°C to -50.0°C)
Resolution - Temperature 0.01°C (-200 to 300°C); 0.1°C (-200 to 400°C)
Transducer Input 4-20mA
Transducer Resolution 1µA
Scanning Speed All channels in less than 1 sec.
Transducer power supply (internal) 20VDC
Operating ambient temperature 0 to 50°C
Dimensions 240 x 140 x 290mm
Weight 4Kg

DOWNLOADS

Documentation and software downloads for 825 Series Data Logger
Data Sheet 825 Series Data Logger

ACCESSORIES

Recommended accessories for 825 Series Data Logger
Product description Code
Carrying Case ECC
P600 Refrence Thermometer E600
P795 Refrence Thermometer E795
Anville Series 275 Temperature Calibration bath (Range -20 to +140°C) E275CC
Anville Series 300 Temperature Calibration bath (Range +40 to +425°C) E300CC
Ultra-low Temperature Calibration bath (Range -80 to +100°C) ERCB80
Class 1 Type T Exposed Junction Thermocouple Sensors (specift length) S-TC-xx
Validation Documentaton ETHVD
Anville Thermocouple Entry Fitting ETCEF
Pressure Transducer (Range -0.830 to +4 bar gauge. Accuracy 0.05% terminal point) EPT
Portable Pneumatic Calibrator (Range: -0.950 bar to 4 bar gauge. Accuracy: ±0.05% of total error band) EPPC