Autoclave Button Logger
PRINT E-MAIL BOOKMARK  

SL55TA Autoclave button logger is an ideal tool for monitoring and mapping autoclaves. It is an extremely compact, size of a watch battery, self contained single channel unit that has temperature range of +15°C to +140°C and can record up to 8000 data readings.

A USB interface reader allows the device to be connected to a PC for the data to be downloaded to be further analysed using TempIT-Lite software downloadable from this site. The software can also be up-graded to TempIT-Pro which provides additional features such as providing ability to carry out calculations of MKT, F0, A0 and PU.

It is recommended that the logger be placed in a SS enclosure before placement into the autoclave. Additional accessories available include printer to directly print from logger and brightly colored enclosures, so as not to lose the logger.


  ENQUIRY
MODELS

Product description Code
SL55TA Autoclave button logger ESL55TA

 


MORE INFORMATION

FEATURES

Product Details for Autoclave Button Logger
bullet Accuracy of ±0.5°C from +80°C to +140°C
Productseperator
bullet Ideal tool for monitoring and mapping autoclaves
Productseperator
bullet Battery life of around 300 hours when used above 50°C
Productseperator
bullet User programmable from 2 seconds to 24 hours
Productseperator
bullet Compact size, as small as a watch battery

SPECIFICATIONS

Product specifications for Autoclave Button Logger
Temperature range +15°C to +140°C
Accuracy ±0.5°C from +80°C to +140°C
±1.0°C from +15°C to +80°C
Battery life Typically 300 hours over 50°C
Sample rates (user selectable) 2 seconds to 24 hours*1 & 2
Memory size 8192 - Low Resolution Mode
4096 - High Resolution Mode
Battery 3.0V Lithium: Non Replaceable
Resolution 8 Bit - Low Res Mode (0.5°C)
11 Bit - High Res Mode (0.07°C)
Response time Typically 90 seconds in circulating air
Size 17mm Diameter x 6mm Height
Case material 305 Stainless Steel
Weight 4 grams
Sealing IP55 - Water Resistant See protective enclosures for waterproof applications
Communication RS232 Serial or USB
Clock accuracy ±2 Minutes per Month
Sampling methods Continuous (First in First Out) or Stop When Full*2

DOWNLOADS

Documentation and software downloads for autoclave button logger
Data Sheet Autoclave button logger enclosure broucher

ACCESSORIES

Recommended accessories for SL55TA
Product description Code
Stainless steel enclosure SL50-ACC01
Fast response stainless steel enclosure SL50-ACC05
Rubberised Holder C/W magnet SL50-ACC02
Handheld printer & charger SL50-ACC09